Establishment of regional diagnostic reference levels for digital mammography in Western Province of Sri Lanka

Show simple item record

dc.contributor.author Niroshani, H.S.
dc.contributor.author Jeyasugiththan, Jeyasingam
dc.contributor.author Senanayake, G.
dc.contributor.author Negishi, Toru
dc.date.accessioned 2021-07-09T07:36:29Z
dc.date.available 2021-07-09T07:36:29Z
dc.date.issued 2021
dc.identifier.citation H S Niroshani et al 2021 J. Radiol. Prot. 41 79 en_US
dc.identifier.uri 10.1088/1361-6498/abd075
dc.identifier.uri http://archive.cmb.ac.lk:8080/xmlui/handle/70130/5511
dc.description.abstract The radiation dose to the breasts should be kept to a minimum as breast tissues are highly sensitive to radiation. In mammography, the mean glandular dose (MGD) is used to specify the patient dose. In this study, data on the MGD during diagnostic mammographic examinations was collected using the database from six digital mammography facilities available in the Western Province in Sri Lanka. Examinations involving breast pathology, breast implants, or compressed breast thicknesses (CBT) outside the range of 20–110 mm were excluded in this study. The mean MGD per breast was 3.50 mGy, with a mean CBT of 57 mm. The mean MGD per facility varies from 1.58 to 2.27 mGy, with overall 75th and 95th percentiles of 2.15 and 2.82 mGy, respectively. The 75th and 95th percentile MGD per image, for the average CBT of 57 ± 12 mm, were 2.00 and 2.65 mGy respectively. The 75th percentile value of the MGD is suggested for the Western Province and it depends on the specific CBT. en_US
dc.language.iso en en_US
dc.title Establishment of regional diagnostic reference levels for digital mammography in Western Province of Sri Lanka en_US
dc.type Article en_US