Please use this identifier to cite or link to this item: http://archive.cmb.ac.lk:8080/xmlui/handle/70130/7427
Title: Investigation of 60Co BEBIG High Dose Rate Brachytherapy Dose Distribution in Different Media Using GEANT4 Monte Carlo Simulation
Authors: Jayakody, M.
Jeyasugiththan, J.
Peterson, S.W.
Rajasooriyar, C.
Chougule, A.
Keywords: HDR Brachytherapy
60Co
GEANT4 Simulation
Dose Distribution
Tissue Heterogeneity
Issue Date: 2024
Publisher: University of Colombo
Citation: Jayakody, M., Jeyasugiththan, J., Peterson, S.W., Rajasooriyar, C., and Chougule, A. (2024). Brachytherapy Dose Distribution in Different Media Using GEANT4 Monte Carlo Simulation. Proceedings: University of Colombo Annual Research Symposium 2024, p.302.
Abstract: High Dose Rate (HDR) Brachytherapy (BT) using 60Co source, uses radioactive sources placed inside patient's body, targeting tumour regions. Consequently, treatment planning is the main step in HDR-BT, which relies on the computer algorithm 'Treatment Planning System' (TPS). The current approach to TPS is based on the American Association of Physicists in Medicine (AAPM) Task Group No. 43 (TG-43) formalism. A major ...
URI: http://archive.cmb.ac.lk:8080/xmlui/handle/70130/7427
Appears in Collections:Department of Nuclear Science

Files in This Item:
File Description SizeFormat 
nuc1.pdf364.97 kBAdobe PDFView/Open