Please use this identifier to cite or link to this item: http://archive.cmb.ac.lk:8080/xmlui/handle/70130/5511
Title: Establishment of regional diagnostic reference levels for digital mammography in Western Province of Sri Lanka
Authors: Niroshani, H.S.
Jeyasugiththan, Jeyasingam
Senanayake, G.
Negishi, Toru
Issue Date: 2021
Citation: H S Niroshani et al 2021 J. Radiol. Prot. 41 79
Abstract: The radiation dose to the breasts should be kept to a minimum as breast tissues are highly sensitive to radiation. In mammography, the mean glandular dose (MGD) is used to specify the patient dose. In this study, data on the MGD during diagnostic mammographic examinations was collected using the database from six digital mammography facilities available in the Western Province in Sri Lanka. Examinations involving breast pathology, breast implants, or compressed breast thicknesses (CBT) outside the range of 20–110 mm were excluded in this study. The mean MGD per breast was 3.50 mGy, with a mean CBT of 57 mm. The mean MGD per facility varies from 1.58 to 2.27 mGy, with overall 75th and 95th percentiles of 2.15 and 2.82 mGy, respectively. The 75th and 95th percentile MGD per image, for the average CBT of 57 ± 12 mm, were 2.00 and 2.65 mGy respectively. The 75th percentile value of the MGD is suggested for the Western Province and it depends on the specific CBT.
URI: 10.1088/1361-6498/abd075
http://archive.cmb.ac.lk:8080/xmlui/handle/70130/5511
Appears in Collections:Department of Nuclear Science

Files in This Item:
File Description SizeFormat 
NS3.pdf308.31 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.